My sample only takes 1 drop to reach an end-point.

Sulfite Testing

Sulfite-1

If the sample reaches an end-point after one drop, you cannot tell if there is zero sulfite or a low amount (under the titrant ppm per drop strength). We recommend using a weaker titrant to verify tests, if possible; or double the sample size and divide the drop equivalency in half. Verify the chemical feed equipment, lines and treatment are operational.